Beveiligingsadvies

CVE-2019-25719

CVE-kwetsbaarheidsdetails - eXtreme Datacenter Security Operations

Gepubliceerd 2026-06-02 14:00:52
Laatst bijgewerkt 2026-06-03 15:43:03
Toegewezen door VulnCheck
CVSS-score 8.8
Status PUBLISHED

Beschrijving

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.