Beveiligingsadvies

CVE-2024-9832

CVE-kwetsbaarheidsdetails - eXtreme Datacenter Security Operations

Gepubliceerd 2024-11-14 21:03:16
Laatst bijgewerkt 2024-11-18 15:37:00
Toegewezen door Baxter
CVSS-score 9.3
Status PUBLISHED

Beschrijving

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.